Alarms

When an alarm occurs, the system will stop the therapy session, sound an audible alert and illuminate
in RED the function key of the alarm area.
To stop an alarm, press the function key that is flashing.
Cold Therapy may alarm if:
• Too low or too high a therapy
temperature may be detected
by the system.
Dynamic Compression may
alarm if:
• The time to inflate the pad is
too short or too long.
Sequential Compression may
alarm if:
• The time to inflate the pad is
too short or too long.
Should the system fail to restart after an alarm has sounded and the stated resolutions have not reset
the System, contact Customer Care at Breg, 1.800.321.0607 or 1.760.795.5440 for further assistance.
To resolve:
• Check the tube connection at the
container and at the cold-compression pad.
• Ensure appropriate ice or water level.
• Activate the cold therapy and press start. If
the issue is not resolved, press the cold
therapy button to deactivate.
• If prescribed, ensure the sequential therapy is
activated and press Start/Stop button to
resume sequential therapy. Contact Customer
Care for further instructions.
To resolve:
• Check the tube connection at the container
and at the cold-compression pad.
• Make sure the cold-compression therapy
pads are properly applied to the body and
connected properly.
• Press dynamic compression and then
press the Start/Stop button. If the issue
is not resolved, contact Customer Care for
further instructions.
To resolve:
• Make sure the sequential compression
therapy pads are properly applied to the
calf and connected properly.
• Press sequential compression and then
press the Start/Stop button. If the issue
is not resolved, contact Customer Care for
further instructions.