Opening The Pad Package; Connecting Defibrillation Pads To The Rescue Sam; Applying Pads To The Patient - Progetti RESCUE SAM ELRS-002 Manuel D'utilisation

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

12

4.3.3 Opening the Pad Package

Open the pad package by tearing along the dotted line, starting at the black arrow (follow directions on
the package). Pull the protective backing from the pads and check that the pads are:
• Free from obvious signs of damage.
• Clean of excessive debris (for example, dirt if the pad was dropped).
• Not dried out, and that the gel is sticky and will adhere to the patient.
• Not expired. Do not use pads after the expiration date printed on the package.
If any of these conditions is found, use a new set of pads.

4.3.4 Connecting Defibrillation Pads to the RESCUE SAM

The RESCUE SAM is designed to be stored with the defibrillation pad connector attached to the unit, while
the pads themselves remain sealed in their package. This reduces the time needed to setup and start
treatment in an emergency.
The AED should be stored with the pad connector plugged into the unit. However, if pads were damaged
or not properly connected, you may need to substitute a new set of pads during an emergency. The pad
connector is on the bottom left corner of the AED.
To remove an old set of pads, pull firmly on the pad connector. Do not reuse used pads. Insert the
connector for the new pads as shown. The connector will only fit in one way – if the connector does not
fit, rotate the connector before trying again. Insert connector firmly until it is completely seated in the
unit.

4.3.5 Applying Pads to the Patient

Correct pad placement is essential for effective analysis of the patient's cardiac rhythm and subsequent
shock delivery (if required). Remove the pads from the pad package by tearing the package along the
dotted line near the top of the package. Remove the pads from the package and follow the directions and
diagram showing proper defibrillation pad placement located on the defibrillation pad package. Peel off
the protective backing from each pad before placing it as shown on the picture on the pad. Peel the
backing off only when the pad is ready to be placed. Place the pads with the sticky side of the pad on the
patient's skin. Place the pads as shown in the left side diagram.
On the right side of the diagram are indicated each pad position marking printed on.
Pads position marking
RESCUE SAM User Manual vers.1.0 March 2011
Pads placement

Publicité