15 Place the Charger below the Stimulator.
16 Slowly move the Charger up, towards the Stimulator.
17 Stop moving the Charger when it stops beeping.
18 Make a mark on your skin at the top edge of the Charger, as shown.
The marks on your skin should create a square or crosshair. The center of this crosshair
indicates the center of the Stimulator.
19 Now that you have located the center of the Stimulator, properly align the Charger over the
Stimulator as shown.

Cleaning Instructions

Cleaning the Charger, Base Station, and Power Supply. The components can be cleaned using alcohol or a mild
detergent applied with a cloth or tissue. Residue from soapy detergents should be removed with a damp cloth. Do not use
abrasive cleansers for cleaning. Do not clean the Charger, Base Station, or Power Supply while they are connected to a
power outlet.
Cleaning the Charging Collar. Hand wash the Charging Collar with mild soap and warm water. Do not machine wash
the charging collar. Let the Charging Collar air dry. Be sure to remove the Charger and Counterweight from the Charging
Collar before washing the Charging Collar.