Installation
Multiple hazards. Only qualified personnel must conduct the tasks
described in this section of the document.
Installation guidelines
High internal temperatures will damage the instrument components.
• Do not operate the instrument in direct sunlight or next to a heat
source.
• Install the instrument indoors in a clean environment with minimum
vibration.
Note: Particulate contamination from the surrounding area is a major cause of
error.
• Install the instrument level.
• Keep the flow rate and operating temperature as constant as possible
for best performance.

Mechanical installation

Instrument mounting
Install this instrument on a flat, vertical surface. Refer to the illustrated
steps in
Figure
3. Mounting hardware (two ¼-20 bolts) is supplied by the
customer.
Install the instrument as near to the sampling point as possible to
minimize lag time of the sample flow.
Refer to
Figure 4
for instrument dimensions and clearance
specifications.
Note: As an alternative, this instrument can be installed on the optional floor stand.