Table 1 Interpretation of verification values
Verification values
Result identified
± 5 mNTU of the
The instrument is
2
baseline
calibrated.
- 5 mNTU (or less) of
The instrument is not
2**
the baseline
calibrated.
+ 5 mNTU (or
The instrument body
greater) of the
may be contaminated.
baseline
There may be fluid on
the optical surface of the
dry standard.
*
If the instrument consistently does not meet the PFC, contact the
manufacturer.
**
A significantly low value is 25% or less of the expected value. A significantly
low value may occur because the light source or detector is not working
correctly. If the verification value is consistently significantly low, send the
instrument to the manufacturer for service.
Verification with a dry standard without software
This procedure includes instructions on how to:
• Assign an initial baseline value to the dry standard without dry
standard software
• Do a dry standard verification without dry standard software
Find and store the baseline value for the dry standard immediately after
the instrument is calibrated and after every subsequent calibration. Refer
to
Assign a baseline value to the dry standard
The baseline value is referenced for subsequent calibration verifications
until the next calibration.
Q-Pulse Id VM277
Vendor Manual Hach A090-070427-003 FT660 SC User Manual
Suggested action
None
1. Clean the instrument.
2. Do the verification again.
3. Recalibrate the instrument
if necessary.
4. If recalibrated, assign a
new baseline value.
1. Clean the body of the
instrument, bubble trap,
detector and dry standard.
2. Recalibrate the instrument.
3. Assign a new baseline
value.
on page 15.
Active 09/09/2013
Note: This procedure can be done on instruments that use the AquaTrend
controller by manually recording the baseline value for the dry standard, pass or fail
criteria (PFC) for verification and verification value.
Note: When not in use, keep the dry standard in the protective case.
Assign a baseline value to the dry standard
1. Remove the head assembly from the body of the instrument.
2. Remove the dry standard from the protective case. Refer to
on page 14.
3. Clean, dry and inspect the dry standard.
a. Use the disposable wipes provided to clean the dry standard.
b. Use a dry cloth to dry the top
Figure 8
on page 14.
c. Make sure that the glass of the dry standard is free of chips and
scratches.
4. Install the dry standard.
a. Put the dry standard in the aperture plate. Make sure that the
locking pin goes in the notch in the aperture plate. Refer to
Figure 8
on page 14.
1
b. Turn the dry standard
/
8
locks in position.
c. Use a dry cloth to remove any fingerprints and oils from the glass
of the dry standard.
5. Put the head assembly on the body of the instrument.
6. When the instrument reading is stable, record the value.
This value is one of three values that will be used to calculate an
average value. The average value will be the final baseline value.
7. Prepare the dry standard to be measured again.
a. Remove the head assembly from the body of the instrument.
1
b. Turn the dry standard
/
8
remove the dry standard from the aperture plate.
c. Do steps 3–5.
8. When the instrument reading is stable, record the value.
Figure 8
1
/
of the dry standard. Refer to
3
of a turn clockwise until the dry standard
of a turn counter-clockwise, then
Page 15 of 68
English 15
®