Hach FT660 sc Manuel D'utilisation De Base page 10

Masquer les pouces Voir aussi pour FT660 sc:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Specifications
Specifications are subject to change without notice.
Specification
Details
Range
0.000–5000 milli-Nephelometric Turbidity Units
(mNTU)
(0–5.0 NTU with +10% over-range limit)
Measurement units
mNTU, mFTU, NTU, FTU, mg/L, Degree
*
Over-range limit
+10%
Method compliance
USEPA Approved for drinking water applications by
Method 10133
Lowest expected reading
7 mNTU
Accuracy
±3% of reading or ±5 mNTU (whichever is greater)
Repeatability
±0.42% at 50 mNTU
±0.3% at 800 mNTU
Resolution
0.001 mNTU on lowest range
0.1 mNTU on highest range
Calibration
Single point at 800 mNTU ±50 mNTU
Calibration frequency
Every 3 months (or as per regulations)
Calibration (wet)
Pass/fail criteria or tolerance (standards less than
verification method
1000 mNTU): 25–50 mNTU
Pass/fail criteria or tolerance (standards greater than
1000 mNTU): 5%
Calibration (dry)
Verification level: 20–50 mNTU
verification method
Performance default: ±5 mNTU of the identified
baseline value
Method precision: less than or equal to 5% relative
standard deviation (RSD)
Signal averaging
No averaging, 6, 30, 60 or 90 seconds
Q-Pulse Id VM277
Vendor Manual Hach A090-070427-003 FT660 SC User Manual
Active 09/09/2013
Specification
Details
Sample light source
Class 1 Laser Product with embedded 7.5 mW,
670 nm, Class 3B Laser Source
Complies with 21 CFR 1040.10 except for deviations
pursuant to Laser Notice No. 50, dated June 24,
2007. FDA Laser Accession No. 9911570
Power requirement
10.5 to 13.2 VDC at 1.5 VA
Operating conditions
0 to 40 °C (32 to 100 °F), 5–95% non-condensing
Storage conditions
–20 to 60 °C (–4 to 140 °F)
Sample requirements
Temperature: 0 to 50 °C (32 to 121 °F)
Flow: 100–750 mL/minute (1.6–11.9 gal/hour)
Recorder output
0–20 mA or 4–20 mA
The output span is programmable over any portion of
the 0.000–1000 mNTU range.
Sample inlet fitting
¼-in. NPT female, ¼-in. compression fitting
Drain fitting
½-in. NPT female, ½-in. hose barb
Tubing requirements
Sample inlet: ¼-inch OD rigid or semi-rigid tubing
Drain: ½-inch ID flexible plastic tubing
Dimensions
25.4 x 30.5 x 40.6 cm (10 x 12 x 16 in.)
(W x D x H)
Mounting options
Wall mounted or floor stand
Shipping weight
Turbidimeter and controller—16.31 kg (13.5 lb)
Turbidimeter only—4.71 kg (10 lb)
ASTM conformance
D6698 Standard test method for online measurement
of turbidity below 5 NTU
Limit of detection (LOD)
0.296 mNTU (ISO 15839)
Protection class
III
Pollution degree
2/II
Page 3 of 68
English 3

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Filtertrak ft660 sc

Table des Matières