CAUTION
Be sure to clean and lubricate the handpiece prior to autoclaving. If blood remains on the internal surface it can become clotted and cause
product failure.
Do not autoclave the product with other instruments even when it is in a pouch. This is to prevent possible discoloration and damage to the
product from chemical residue on other instruments.
Do not heat or cool the product too quickly. Rapid change in temperature could cause damage to the product.
To avoid product failure, do not use a sterilizer whose temperature exceeds 138°C during use including the dry cycle. The chamber
temperature of some type of sterilizers may exceed 138°C. Contact the sterilizer manufacturer for detailed information.
Keep the product in suitable atmospheric pressure, temperature, humidity, ventilation, and sunlight. The air should be free from dust, salt and
sulphur.
Do not touch the product immediately after autoclaving as it will be very hot and must remain in a sterile condition.
Autoclave sterilization is recommended for the product. The validity of other sterilization methods is not confirmed.
NOTICE
NSK recommends Class B sterilizers as stated in EN 13060.