7. Checking the Handpiece Before Each Use
Follow the check below before use. If any abnormalities are found, stop using the handpiece immediately
and contact your Authorized NSK Dealer.
1) Check the Head Cap is firmly tightened.
2) Check the coolant water is flowing properly.
3) Mount the bur. (Refer to "5. Mounting and Removing the Bur")
4) Rotate the handpiece for about one minute with coolant water at the Max. rotation speed of the attached
motor. During rotation, check for abnormalities such as abnormal rotation, vibration, noise.
5) After the handpiece rotation has completely stopped, touch the handpiece head to confirm the head is
NOT heating abnormally (Fig. 5).
CAUTION
To avoid injury, keep your hands away from the bur during rotation.
14
Fig. 5