NSK automatic handpiece cleaning and lubrication system
When using NSK automatic handpiece cleaning and lubrication system, refer to the system instructions.
WARNING
8-6 Sterilization
Sterilize the product by autoclave sterilization. After each patient sterilize as noted below.
1) Insert into an autoclave pouch. Seal the pouch.
2) Autoclavable under the conditions below.
- Gravity displacement steam sterilizer; Exposure at 132 ºC for 15 minutes, Drying for 30 minutes/Exposure at 135 ºC for 10 minutes, Drying for
30 minutes
- Dynamic air removal steam sterilizer: Exposure at 132 ºC for 4 minutes, Drying for 30 minutes/Exposure at 135 ºC for 3 minutes, Drying for
16minutes
3) The handpiece should remain in the autoclave pouch until required for use.
22