www.oped.de | www.oped.ch
Useful details
Technical data
MDD class
IIa (93/42/EEC, Annex IX, Rule 9)
Protection class
II, Type BF
CE certification No.
CE 0123
Rating
Input:
230 V
Power supply:
Manufactured by
Mean Well Enterprises Co., Ltd
Power supply model:
PD-25B
Fuses:
2xT1A250V
DIMDI No.
10-969 in accordance with UMDNS
Classification
UMDNS
compression device
intermittent
Code of the
authority responsible
DE CA 57 (Upper Bavaria)
Article-Nr.
VXSY-06-1
This device is intended for continuous use. EMC tested in accordance
with IEC 60601-1, IEC 60601-1-2 and IEC 60601-1-11.
For information about potential electromagnetic interference and
advice on how to avoid or minimize such interference, please refer
to the attachment of this user manual.
Environmental conditions for transport/storage
• Temperature range: -20° C to 60° C
• Relative humidity: up to 95 % without condensation
• Ambient pressure: 700 hPa to 1.060 hPa
Environmental conditions during operation
• Temperature range: 10° C to 40° C
• Relative humidity: up to 95 % without condensation
• Ambient pressure: 700 hPa to 1.060 hPa
Device dimensions
Height:
23 cm, overall
Depth without handle:
15.5 cm
Width:
35 cm
Weight:
4.3 kg
Warranty
For the VADOplex system, a warranty period of 1 year after delivery
to the patient or distributor is granted.
0123
, 50 Hz, 0.4 A
Useful details
Conversion table
for pressure units mmHg/kPa
Program 1 (Foot)
130 mmHg = 17,3 kPa, 1 Sec. Dwell time, 20 Sec. cycletime
Program 2 (Calf)
130 mmHg = 17,3 kPa, 3 Sec. Dwell time, 50 Sec. cycletime
Program 3 (Hand)
80 mmHg = 10,7 kPa, 1 Sec. Dwell time, 20 Sec. cycletime
Risk of falling
Risk of explosion
Before opening the device, unplug the power supply.
The device meets all essential requirements of the applicable
EU guidelines.
0123
Type BF applied part.
Observe accompanying documents!
The product and its packaging may not be disposed as household waste,
but has to be sent back to the manufacturer for disposal.
Protection class II
Manufacturer
Date of manufacture
40°C
Temperature range
10°C
Data-Matrix
Safety Note
Safety against contact by fingers or
foreign objects langer than 12,5 mm o
Safety against condensation
Single use only (in case of pads)
51