www.oped.de | www.oped.ch
Routine Maintenance
Interval
Routine Maintenance
As required
Thoroughly clean external surfaces of the
controller before transfer between patients
and after prolonged periods of storage.
Attention should be paid to the fan grills,
ensure these are clean at all times.
Annually/
Inspect the AC powersupply plug and cable
3.000 hours
for damage. If this display appears, please
send back the VADOplex System to the
manufacturer for service.
Application note
Do not expose the VADOplex device to extreme temperatures, water
or humidity. Do not drop the device and do not use force trying
to open it. Should one of these events happen, have the device
inspected by an authorized service technician.
Opening the housing or exposing the device to humidity/dampness
creates a risk of electric shock.
Using the device near inflammable gases can cause explosive
incidents. Do not place the device in the vicinity of such gases.
The device may only be used with the pads intended.
The pads have to be disposed of properly.
Attaching the VADOplex at the patient bed
Note: If located on the floor take care not to crush the controller
when altering the bed height.
Cleaning, maintenance and technical service
Only employees, who have undergone the corresponding technical
training from the manufacture are authorized to open the device and
perform maintenance tasks.
The VADOplex system has to be maintained every 3000 operating
hours or at least once a year. During maintenance, the hardware and
software of the device is inspected, wear parts are checked and, if
applicable, exchanged or cleaned. This inspection requires special
training to ensure smooth operation of the VADOplex systems.
Thus, maintenance tasks and technical procedures at the VADOplex
system may only be performed by authorized technicians. Otherwise,
the manufacturer does not assume any responsibility for the use and
effect of the VADOplex system.
The case of the device and the connection hoses can be wipe-
disinfected with standard disinfectant agents for hospital use or
alternatively with a 1% sodium hypochlorite solution.
Residues of alcohol and sodium hypochlorite, remaining on the
surfaces after wipe disinfection, must be removed with a water
moistured cloth.
The user/patient does not have to clean the device. This is done by
the service staff during maintenance/repair.
The VADOplex system does not have to be sterilized.
On request, the manufacturer can supply circuit diagrams, spare
parts lists and other documents for use only by OPED certificated
and qualified expert staff!
49