6/ MAINTENANCE

1
WINTERIZING
In case of pool winterizing (draining and/or stop of the filtration), achieve the following operations :
✓ Remove the pH and redox probes of the pipe (if installed).
✓ Keep it in his original protection filled with city water (or in a glass), in a place where it will not freeze
during winter.
✓ On the pipe, instead of the probe, use the supplied plug.
✓ For the pH regulation, rinse the flexible pipe of the peristaltic pump with clean water. Put a tin of water
in the place of the pH corrective agent tin and use the function "MA" of the pH control menu (see §4.2,
p.23-24).
The pH and redox measure probes are submitted to a reference curve which according to the use of the
pool moves in the time. For an optimal regulation, it is important to proceed to a calibration at each pool
start up, change of water or change of probe.

CALIBRATION OF THE REDOX PROBE

2
2
Before proceeding to a calibration, provide you of the standard solution (470 mV).
e
ct
Unscrew then remove the mobile nut of the probe holder while leaving the probe inside. Use the cork
ur
supplied to fill the opening.
e
✓ Reach the setting menu : press 3 s on the key
d
u
✓ Select the code « CA » with the keys
p
The screen flashes « 47 », the value of the calibration.
H
p
Modèle Uno :
C A
ar
di
Tremper la dans une solution étalon à 470 mV et valider par
o
The screen will flash alternatively 47 and – – during the calibration (1 to 5 minutes).
d
e
4 7
- -
1
If the measured value is right, the device shows « oh » and then return to the previous menu.
If one of the values is wrong, the device shows « Er ». Your probe is probably no more usable (lifespan),
contact your professional to proceed to the replacement.
✓ Rinse the redox probe in water and drain it without drying it with a rag.
✓ Remove the cork and then put back the probe on the probe holder.