Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 28
4. Dry device with a lint-free single-use cloth.
SmartLite
Focus
®
4.3.3 Autoclaving (optional)
The SmartLite
1. Steam autoclave SmartLite
3 minutes 30 seconds.
SmartLite
Focus
®
5

Maintenance

5.1 Light output monitoring
1. Make sure that the LED aperture is clean and scratch-free; otherwise light output
will be reduced and may be insufficient for proper curing of the material.
2. The light intensity of SmartLite
appropriate curing by using a commercially available radiometer. As radiometers
differ greatly, the following procedure is recommended.
3. Upon receipt of SmartLite
your choice and record the value. This initial reading is for reference only, and may
not represent the actual light intensity.
4. For subsequent monitoring retest the light intensity with the same radiometer that
was used for the initial reference and compare it to the reference intensity.
5. If radiometer value drops below reference intensity, light efficiency can be verified
using the i•Cure. Place i•Cure on a sheet of paper on a flat surface. Choose
i•Cure segment according to step height required (please note that the step
height should be twice the curing depth to be confirmed). Fill with composite. Hold
SmartLite
Focus
®
on the lower aperture has been cured (i.e. cannot be scraped off with a plastic
spatula), the curing depth according to ISO 4049:2009 equals half the chosen step
height. If i•Cure test is successful, SmartLite
6. Do not continue to use SmartLite
intensity and i•Cure test has failed.
5.2 Battery
2
• Battery is equipped with low self discharge technology resulting in a long operating
life.
• Battery is pre-charged and ready to use upon purchase. Charging before first use
may improve initial operating time but is not mandatory.
• When the battery light flashes green
recharging, the battery light remains permanently green. The battery needs ap-
proximately 3 hours to be fully recharged. SmartLite
ing Technology. Depending on battery status, SmartLite
choose between Automatic Quick Charging Mode (fast flashing green light) and
Standard Charging Mode (flashing green light).
• The SmartLite
5 minutes operating time after 10 minutes recharging.
• When the battery light flashes yellow (fast)
charged. At first occurrence approximately 1 minute curing time remain for comple-
tion of the treatment. Light output is not reduced during this period. After this period,
the device switches off automatically and the battery light remains permanently
yellow indicating the need to recharge.
Shields can be disinfected for up to 100 times.
®
Focus
Shields can also be autoclaved.
®
®
Focus
®
Shields can be autoclaved for up to 60 times.
®
Focus
®
in close proximity to the upper aperture and cure. If the material
®
Focus
Automatic Quick Charging Mode allows a minimum of
®
®
Shields at 134 °C/2 bar = 273 °F/29 psi for
®
Focus
should be checked frequently to ensure
®
®
check the light intensity with the radiometer of
®
Focus
®
Focus
if both light output is below reference
®
®
, the battery is charging. Upon complete
1.2
1.3
is ready to use.
®
Focus
has a Smart Recharg-
®
®
Focus
will automatically
®
®
, the battery needs to be re-
9

Publicité

Table des Matières
loading

Table des Matières