User Settings - OHAUS Valor 1000W Serie Guide De L'utilisateur

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

TM
OHAUS Valor
1000W Series

4.2 User Settings

Press and hold SET, then
user settings [F1-0 x]:
Parameters
Description
[F1-0 x]
Key Zero Range
[F1-1 x]
Auto-zero Tracking Range
[F1-2 x]
Initial Zero Range
[F1-3 x]
Unit of Measure
[F1-4 x]
Filter Level
[F1-5 x]
Stability Indication
[F1-6 x]
Display Brightness
[F1-7 x]
Accumulation Zero Band
(M+ is effective when stable
weight returns to within "x")
Press
to step through available settings.
Press YES to confirm displayed setting and proceed to next menu item.
Press SET at any time to save and exit.
until
[SEt-F1] is displayed. Press YES to enter and view
Settings of "x"
0 = 2% of scale capacity (default)
1 = 5% of scale capacity
2 = 10% of scale capacity
3 = 20% of scale capacity
0 = 0ff
1 = 0.5d (d = scale division) (default)
2 = 1d
3 = 2d
4 = 3d
5 = 4d
6 = 5d
0 = 10% of scale capacity (default)
1 = 20% of scale capacity
2 = 50% of scale capacity
3 = 100% of scale capacity
0 = kg (default)
1 = lb
0 = Low
1 = Medium (default)
2 = High
3 = maximum
0 = Fast
1 = High (default)
2 = Medium
3 = Slow
0 = Low
1 = Medium (default)
2 = High
3 = maximum
0 = 0d (d = scale division) (default)
1 = within 1d
2 = within 2d
3 = within 5d
4 = within 10d
EN-7

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Valor v11pw3Valor v11pw6Valor v11pw15Valor v11pw30

Table des Matières