portable and mobile RF communications equipment and the JFT90
The JFT90 is intended for use in an electromagnetic environment in which radiated RF disturbances
are controlled. The customer or the user of the JFT90 can help prevent electromagnetic interference by
maintaining a minimum distance between portable and mobile RF communications equipment (trans-
mitters) and the JFT90 as recommended below, according to the maximum output power of the com-
munications equipment.
Rated maximum output
power of transmitter
W
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation
distance d in metres (m) can be estimated using the equation applicable to the frequency of the trans-
mitter, where P is the maximum output power rating of the transmitter in watts (W) according to the
transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
Recommended separation distances between
Separation distance according to frequency of transmitter
150 kHz to 80 MHz
d = 1.2 P
0.12
0.38
1.2
3.8
12
Non-contact fever thermometer JFT 90
m
80 MHz to 800 MHz
d = 1.2 P
0.12
0.38
1.2
3.8
12
GB
800 MHz to 2,5 GHz
d = 2.3 P
0.23
0.73
2.3
7.3
23