Guidance and manufacturer's declaration - electromagnetic immunity
The JFT90 is intended for use in the electromagnetic environment specified below. The customer or the
user of the JFT90 should assure that it is used in such an environment.
Immunity test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast transient/
burst
IEC 61000-4-4
Surge
IEC 61000-4-5
IEC 60601 test level
6 kV contact
8 kV air
2 kV for power supply
lines
1 kV for input/output lines
1 kV line(s) to line(s)
2 kV line(s) to earth
Non-contact fever thermometer JFT 90
Compliance level
Floors should be wood,
concrete or ceramic tile.
6 kV contact
If floors are covered with
8 kV air
synthetic material, the
relative humidity should
be at least 30 %.
Mains power quality
should be that of a typi-
Not applicable
cal commercial or
hospital environment.
Mains power quality
should be that of a typi-
Not applicable
cal commercial or
hospital environment.
GB
Electromagnetic
environment -
guidance
137