Guidance and manufacturer's declaration – electromagnetic immunity
The JFT90 is intended for use in the electromagnetic environment specified below. The customer or
the user of the JFT90 should assure that it is used in such an environment.
Immunity
IEC 60601 test
test
level
Conducted
3 Vrms
RF
150 kHz to 80
IEC 61000-
MHz
4-6
Radiated RF
3 V/m
IEC 61000-
80 MHz to 2.5
4-3
GHz
Non-contact fever thermometer JFT 90
Compliance
Electromagnetic environment – guidance
level
Portable and mobile RF communications equip-
ment should be used no closer to any part of the
JFT90, including cables, than the recommended
separation distance calculated from the equation
applicable to the frequency of the transmitter.
Recommended separation distance
d = 1.2 P
Not applicable
d = 1.2 P 80 MHz to 800 MHz
d = 2.3 P 800 MHz to 2.5 GHz
where P is the maximum output power rating
of the transmitter in watts (W) according to the
transmitter manufacturer and d is the recommen-
3 V/m
ded separation
Distance in metres (m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey,
should be less than the compliance level in each
frequency range.
Interference may occur in the vicinity of equip-
ment marked with the following symbol: