Electromagnetic compatibility

Electromagnetic emissions
Guidance and manufacturer's declaration – electromagnetic emissions for all EQUIPMENT and SYSTEMS
The Blood pressure monitor (SBM 37) is intended for use in the electromagnetic environment specified below. The customer of the
user of the Blood pressure monitor (SBM 37) should assure that it is used in such and environment.
Emission test
Compliance
RF emissions
Group 1
CISPR 11
RF emission
Class B
CISPR 11
Electromagnetic immunity
Guidance and manufacturer's declaration – electromagnetic immunity – for all EQUIPMENT and SYSTEMS
The Blood pressure monitor (SBM 37) is intended for use in the electromagnetic environment specified below. The customer of the
user of the Blood pressure monitor (SBM 37) should assure that it is used in such and environment.
Immunity test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electromagnetic environment – guidance
The Blood pressure monitor (SBM 37) uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to cause any interference
in nearby electronic equipment.
The Blood pressure monitor (SBM 37) is suitable for use in all establishments, inclu-
ding
domestic establishments and those directly connected to the public low-voltage
power supply network that supplies buildings used for domestic purposes.
IEC 60601
Compliance
test level
level
±6 kV contact
±6 kV contact
±8 kV air
±8 kV air
Electromagnetic environment – guidance
Floors should be wood, concrete or ceramic tile. If floor are
covered with synthetic material, the relative humidity should be
at least 30%.