16
6. MaIntenanCe
6 .1 . HAndPIEcE MAIntEnAncE
• As a routine: after treatment with water flow, flush the water circuits of the handpiece
and tip with water at a low power level for around 20- 30 seconds by pressing down
the foot pedal.
• In case of damage, excessive wear at least once a year: replace the watertight O-ring
which is located in the rear part by using the fitting tool (for details see Fig. 6).
F
Note
Lubricate the water-tight O-ring with
appropriate dental lubricant in order
to prolong its efficacy and prevent
internal leaks.
Fig. 6 Handpiece Maintenance
6 .2 . tIP MAIntEnAncE
Basically, shape and mass of a tip are the determining factors to obtain consistent
performance of the ultrasonic generator.
• The thinner the diameter of a tip, the less the tip should be re-used.
• After a certain time of usage the shape of a tip will gradually be modified, i.e. the tip
will become shorter.
• Operation with a worn tip will lead to longer preparation time and may increase
temperature in the cavity or root canal. It may also lead to tip breakage.
1. Do not actively modify a tip by filing.
2. Regularly check a tip for visible wear with a loup or microscope.
3. Replace
– the tip which is 3 to 4mm short after several operations, compared to the length
of a new tip.
– the tip which has apparently suffered an accidental impact (fall etc.).
– the diamond coated tip as soon as it is worn.
– tips that are bent to match canal curvature cannot be reused.
– follow the use instructions received with authorized replacement tips.
6 .3 . controL unIt MAIntEnAncE
1. Regularly inspect accessories and cords for insulation defects and replace if necessary.
2. Regularly check whether the air vents of the control unit are clean to avoid overheating.
3. It is necessary to remove any traces of moisture in the handpiece connection; if yes
dry it by wiping and airblowing.
WarNiNG
Do not use the device if it appears to be defective.
In case of any malfunction, contact the local supplier or directly DENTSPLY Tulsa Dental
Specialties rather than have it repaired by an unauthorized person.
6 .4 . WAtEr FILtEr rEPLAcEMEnt And MAIntEnAncE
Regularly clean and change the water filter if necessary.
Follow the instructions below:
1. Turn-off the water supply and switch-off
the device (OFF position), then unplug
the power cord.
2. Unscrew the two parts of the filter using
two open-ended 10 mm wrenches.
3. Remove the filter cartridge to replace or
clean it under running water.
4. Proceed in the same manner for the seal.
5. Carry out the procedure in reverse to
reassemble the components.
6. Check that the spray operates correctly
and that there are no leaks.
WarNiNG
Replace a worn or damaged filter cartridge.
17
Fig. 7 Disassembly of Filter
Fig. 8 Reassembly of Filter