7. tRoubleshootInG

If the device does not operate correctly review the checklist below in order to exclude any
user error before contacting your local supplier or DENTSPLY Tulsa Dental Specialties directly.
PossIble Cause
PRobleM
Defective power cord
No operation
connection
Blown fuse
Defective water nose connection
No water spray
No water pressure
Clogged filter
Defective solenoid valve
Obstructed tip or irrigation file
No water at tip but
vibration present
Wrong tip selection
Incorrect spray adjustment
Worn or deformed tip
insufficient power
Presence of fluid or moisture
low vibration
between handpiece and cord
Incorrect tip tightening
No ultrasonic output
Defective connector contact
Cut wire in handpiece cord
Worn O-ring in handpiece
Water leak between
handpiece and its base
between
or at junction
handpiece or cord
F
Note
Find additional information in the DFU of each ProUltra tip.
solutIon
Contact DENTSPLY Tulsa Dental
Specialties for after sales service
Contact DENTSPLY Tulsa Dental
Specialties for after sales service
Check water supply
Check pressure of mains
Clean or replace filter
Contact DENTSPLY Tulsa Dental
Specialties for after sales service
Remove obstruction for tips or
irrigation files
Check tip
Re-adjust water spray
Replace tip
Thoroughly dry electrical contacts
with air
Re-tighten with wrench
Clean cord contacts
Contact DENTSPLY Tulsa Dental
Specialties for after sales service
for cord replacement
Replace the O-ring
8. teChnICal / PeRFoRManCe Data
ProUltra Piezo Ultrasonic
Model/type
Minimum frequency of
28 KHz
vibrations
Control unit: 0.9 kg
Weights
Handpiece: 52 gr
Control unit: 11.2cm x 11.5cm x 12cm
Dimensions (W x D x H)
Foot pedal: 7cm x 9.5cm x 3cm
30 VA
Maximum power
115 VAC
Mains power supply
50 Hz / 60 Hz
Voltage of 100 VAC and 115 VAC: values of fuses 1 and 2
Protection
are 500 mAT (not accessible)
Protection Class II
Device protection class
BF
applied component type
5 mins. Operation / 10 mins. Pause
intermittent operation
1 to 5 bar (14 to 72 PSI)
Water pressure at inlet
60 μm
Water filter
0-90 ml/min
turn half way to adjust
water flow

aMbIent ConDItIons

+10°C/50°F to +40°C/104°F at a maximum of 95% relative
operating temperature
humidity without condensation
-20°C/-4°F to +70°C/158°F at a maximum of 95% relative
storage temperature
humidity without condensation
F
Note
The voltage is preset. The device is protected by fuses (unaccessible for user).