10 Technical Data; Operation And System Construction; Input Values - Endress+Hauser iTEMP TMT184 PROFIBUS-PA Manuel De Mise En Service

Transmetteur de température
Table des Matières

Publicité

10 Technical Data

Measurement principle
Measurement system
Measurement value
Measurement range
Type of input
78
10
Technical Data
10.0.1

Operation and system construction

Electronic measurement and conversion of input signals in industrial temperature measurement.
®
The iTEMP
PA TMT184 temperature head transmitter is a 2-wire transmitter with measurement
inputs for resistance thermometers and resistance transmitters in 2-, 3- or 4-wire connection, ther-
mocouples and voltage transmitters. Applications are in the measurement and control areas for pro-
cess monitoring. The TMT184 Setup is done using the PROFIBUS-PA
PC software (e.g. Commuwin II).
More detailed information for project planning can be found in the operating manual
See »Further documentation« on page 81.
10.0.2

Input values

Temperature (linear temperature transmission), resistance and voltage
The transmitter measures various measurement ranges dependent on sensor connection and input
signal.
Pt100
Pt500
Pt1000
to IEC 751
Ni100
Resistance thermometer
Ni500
(RTD)
Ni1000
to DIN 43760
• Connection modes: 2-, 3- or 4-wire connection
• On 2-wire connection software compensation of the cable resistance is possible (0 to 30
• On 3- and 4-wire connection sensor cable resistance up to max. 11 Ω per core
• Sensor current: ≤ 0.2 mA
Resistance (Ω)
Resistance transmitter
B (PtRh30-PtRh6)
C (W5Re-W26Re)
D (W3Re-W25Re)
E (NiCr-CuNi)
J (Fe-CuNi)
K (NiCr-Ni)
L (Fe-CuNi)
N (NiCrSi-NiSi)
Thermocouple (TC)
R (PtRh13-Pt)
S (PtRh10-Pt)
T (Cu-CuNi)
U (Cu-CuNi)
to IEC 584 Part 1
• Cold junction compensation: internal (Pt100)
• Cold junction compensation accuracy: ± 1 K
Millivolt transmitter (mV)
Voltage transmitter (mV)
Type
-200 to 850 °C (-328 to 1562 °F)
-200 to 250 °C (-328 to 482 °F)
-200 to 250 °C (-328 to 482 °F)
-60 to 250 °C (-76 to 482 °F)
-60 to 150 °C (-76 to 302 °F)
-60 to 150 °C (-76 to 302 °F)
Ω)
10 to 400 Ω
10 to 2000 Ω
1
I
-270 to +1000 °C (-454 to 1832 °F)
-210 to +1200 °C (-346 to 2192 °F)
-270 to +1372 °C (-454 to 2501 °F)
2
-200 to +900 °C (-328 to 1652 °F)
-270 to +1300 °C (-454 to 2372 °F)
-50 to +1768 °C (-58 to 3214 °F)
-50 to +1768 °C (-58 to 3214 °F)
-270 to +400 °C (-454 to 752 °F)
2
-200 to +600 °C (-328 to 1112 °F)
-10 to 75 mV
®
protocol combined with a
Measurement range
0 to +1820 °C (32 to 3308 °F)
0 to +2320 °C (32 to 4208 °F)
0 to +2495 °C (32 to 4523 °F)
TMT184
BA198F/00/en.