On start up the system calibrates the dimmer plate position. If the process fails this
screen will display and the lamp will not ignite. Turn off and restart the illuminator. If the
fault persists the unit needs servicing.
6.2.1.8
Lamp will not strike or stay lit. Replace the lamp. If the fault persists the unit needs
servicing.
6.2.1.9
This screen is displayed when the control system does not get a response from the lamp
during startup. Turn off and restart the illuminator. If the fault persists the unit needs
servicing.
LIT-219 Sunoptic Surgical
Rev. C
DIMMER CALIBRATION ERROR
LAMP FAULT
LAMP COMMUNICATION FAULT
®