The [MDS3001/MDS3001PLUS/MDS3001U] is intended for use in the electromagnetic environment
specified below. The customer or the user of the [MDS3001/MDS3001PLUS/MDS3001U] should assure
that it is used in such an environment.
Phenomenon
Electrical fast transients/burst
Surges
Line-to-line
Surges
Line-to-ground
Conducted disturbances
induced by RF fields
Basic EMC
standard
IEC 61000-4-4
IEC 61000-4-5
IEC 61000-4-5
IEC 61000-4-6
-37-
Immunity test levels
Home Healthcare Environment
±2 kV
100kHz repetition frequency
±0.5 kV, ±1 kV
±0.5 kV, ±1 kV, ±2 kV
3V, 0.15MHz-80MHz
6V in ISM and amateur radio bands
between 0.15MHz and 80MHz
80%AM at 1kHz