then the time and date is confirmed.
d. The monitor will turn off automatically after 1 minute of no operation.
e. Once you change the batteries, you should readjust the time and date.

3. TECHNICAL ALARM DESCRIPTION

The monitor will show 'HI' or 'Lo' as technical alarm on LCD with no delay if the
determined blood pressure (systolic or diastolic) is outside the rated range specified
in part SPECIFICATIONS. In this case, you should consult a physician or check if your
operation violated the instructions.
The technical alarm condition (outside the rated range) is preset and cannot be
adjusted or inactivated. This alarm condition is assigned as low priority according to
IEC 60601-1-8.
The technical alarm is non-latching and does not require a reset. The signal displayed
on LCD will disappear automatically after about 8 seconds.

4. CONNECTING THE CUFF TO THE MONITOR

Insert the cuff tubing connector into the socket in the
left side of the monitor. Make certain that the connector
-17-
Picture 3