The PTFE insulation and emitter pin are clean.
3 Check that no ethyl alcohol is left on the ionizer.
Cleaning
Thoroughly clean the ionizer every three months.
1 Deinstall the ionizer.
2 Dampen the lint-free cloth or foam cotton swab with a small amount of ethyl alcohol.
3 NOTICE: Damage or malfunction Do not touch the emitter pin with any hard object.
NOTICE: Malfunction due to contamination Avoid touching the emitter pin with fingers. Wear gloves.
Wipe the inside and outside of the insulation (white PTFE) around the emitter pin (1). Very gently wipe the
emitter pin.
The PTFE insulation and emitter pin are clean.
4 Clean the insulation tube (3) between the steel emitter barrel (2) and the body of the ionizer.
5 Check that no ethyl alcohol is left on the ionizer.
6 Reinstall the ionizer, see [Installing Antistatic Kit   Page 6].
Contact your METTLER TOLEDO service representative for details about available service options.

7 Technical Data

7.1 General data

Use only with a tested AC Adapter with SELV output current.
Ensure correct polarity
Ionizer
Input voltage:
Input corrent (operating):
Weight:
Protection and standards
Overvoltage category:
Degree of pollution:
Range of application:
Environmental conditions
Height above mean sea level:
Ambient temperature:
Relative air humidity:
Atmosphere:
Ozone
All HV ionizers produce a trace of ozone. The METTLER TOLEDO ionizer ozone level is considerably below the
international safety limit of 0.1 ppm.
10
CAUTION
9 V  to 15 V DC
560 mA
150 g (complete with packaging)
II
2
For use only in closed interior rooms
up to 4000 m
5-40 °C
Max. 80% up to 31 °C, linearly decreasing to 50% at 40 °C,
noncondensing
Do not use under inert gas atmosphere
Compact Antistatic Kit