Error displayed
Error when calling up the local IP address.
Error during default gateway setup.
Error during network drive setup!
Error during subnet mask setup.
Error copying from USB memory stick.
Error in FTP connection.
Fault
Test program stopped!
Please check lamp
Close the lid.
Error [xx]
Fault
Test program stopped!
Please remove the cuvette
Close the lid.
Error
Selfcheck stopped.
Hardware error.
Error [x]
Error
Too much ambient light!
Move instrument into shade
or close the lid
No instrument backup present!
No valid data for these parameters!
Definition
Network setup: DHCP client has no connection to the
DHCP server
Network setup: default gateway cannot be set for fixed
IP address
Error during network setup
Network setup: Subnet mask cannot be set for fixed IP
address
Error during update
FTP error
Test program stops when the instrument is started
Test program stops when the instrument is started
Electronic defect
The instrument sensors detects too much ambient light.
Data analysis not possible, no measurement data
Resolution
Enter the IP address again.
Try to create the connection again.
Check the settings.
Enter the subnet mask again.
Restart the process or contact the manufacturer or
national agent.
Make sure that the instrument is connected to the
network.
Check the lamp and replace it if necessary.
Close lid.
Press Start Again.
Remove the cuvette/sample cell from the cell
compartment.
Press OK.
Contact the manufacturer or national agent and specify
the error number.
Decrease ambient light. Avoid direct sunlight.
Close lid.
Check the USB memory stick.
Change the selection.