Technical Information
Use
Front / Remote Selection
Press CYCLE NUMBER and PUMP CURRENT pushbuttons together
for at least 2 seconds and the processor enters in a routine where it
is possible to program the controller. In this routine, the CYCLE
push-button is used for choosing/changing the value or condition;the
PUMP CURRENT push-button is used to enter and confirm the value.
At any time it is possible to exit this routine by pressing the CYCLE
and PUMP CURRENT pushbuttons at the same time for at least 2
seconds.
The display shows:
C
S
and then
S
S
where: XXX = YES or NO.
If YES is selected, the Soft Start mode allows the pump to ramp-up
the Normal speed within ten steps. When NO is selected, the Soft
Start mode is deselected and the ramp-up of the pump will be done
within 300 seconds.
The controller factory default is set to YES.
The Soft Start mode may be deselected/selected only when the pump is
NOTE
stopped.