6. Changing between Body and Object Mode
1. Press the START/IO button 6. The display 2 is activated to
show all segments for 1 second.
2. The default mode is body mode AL. Press the MODE button 4
to switch to object mode AM. For switching back to body mode,
press the MODE button again.
7. Directions for Use
Always remove the protective cap CT before use.
Measuring in body mode
1. Press the START/IO button 6. The display 2 is activated to
show all segments for 1 second.
2. When the «°C» or «°F» icon is flashing, a beep sound is heard
and the thermometer is ready for measuring 9.
3. The probe LED light is activated and will keep flashing.
4. Straighten the ear canal by pulling the ear up and back to give
a clear view of the eardrum.
For children under 1 year; pull the ear straight back.
For children of 1 year to adult; pull the ear up and back.
Also refer to the short instructions at the front.
5. While gently pulling the ear, insert the probe snugly into the ear
canal.
6. The probe LED light will stop flashing (stays illuminated) and
«good» will be displayed on the LCD, when the measuring
sensor detects an appropriate position.
7. Immediately press the START/IO button 6. Release the button
and wait for the beep sound. This indicates the end of the
measurement.
8. Remove the thermometer from the ear canal. The display shows
the measured temperature AK.
9. For the next measurement wait until the «°C»/«°F» icon is
flashing and follow steps 5-7 above.
10.Press and hold the START/IO button 6 for 3 seconds to turn
off the device; otherwise the device will automatically switch off
after approx. 60 seconds.
Measuring in object mode
1. Press the START/IO button 6. The display 2 is activated to
show all segments for 1 second.
2. Press the MODE button 4 to switch to object mode.
3. Aim the thermometer at the center of the object you want to
measure with a distance of no more than 5 cm. Press the
IR 200
START/IO button 6. After 1 second a long beep will verify the
completion of measurement.
4. Read the recorded temperature from the LCD display.
5. For the next measurement wait until the «°C»/«°F» icon is
flashing and follow steps 3-4 above.
NOTE:
Patients and thermometer should stay in similar room
condition for at least 30 minutes.
To ensure accurate readings, wait at least 30 sec. after 3-5
continuous measurements.
Accumulation of ear wax on the probe can result in less accu-
rate temperature readings or cross infection between users.
It is essential that the probe is cleaned after every measure-
ment. Therefore, this device reminds the user to clean the probe
when turning off the device. «CLEAN ME» BN is displayed and
the probe LED light will flash for 3 seconds. For cleaning, follow
the instructions in the «Cleaning and Disinfecting» section.
After cleaning the measuring sensor 1 with alcohol, wait
5 minutes before taking the next measurement, in order to
allow the thermometer to reach its operating reference temper-
ature.
10 short beeps and a red LCD backlight alert the patient that he/
she may have a temperature equal to or higher than 37.5 °C.
For an infant, it is best to have the child lying flat with his/her head
sideways so the ear is facing upwards. For an older child or adult,
it is best to stand behind and slightly to the side of the patient.
Always take the temperature in the same ear, since the temper-
ature readings may be different from ear to ear.
In the following situations it is recommended that three temper-
atures in the same ear be taken and the highest one taken as
the reading:
1. New born infants in the first 100 days.
2. Children under three years of age with a compromised
immune system and for whom the presence or absence of
fever is critical.
3. When the user is learning how to use the thermometer for
the first time until he/she has familiarized himself/herself with
the device and obtains consistent readings.
4. If the measurement is surprisingly low.
Don't take a measurement while or immediately after nursing a
baby.
Don't use the thermometer in high humidity environments.