Z E R O O K
P R E S S " T E S T "
R U N T E S T
Test
C y s :
5 6
m g / l
Accuracy of the method
The method on hand was developed from a gravimetric process for the determination of cyanuric
acid. Based upon undefined edge conditions the deviations of the standard method may be
greater.
*
PC 01
C
Cyanuric acid *
After zeroing, the display shows:
1.
Remove the cell from the cell compartment.
2.
Add a CYANURIC ACID-tablet to the 10 ml water sam-
ple and crush and mix well with a clean stirring rod.
Fit cell with a clean stopper.
3.
Place the cell in the cell compartment as soon as all
tablet ingredients have dissolved and close the lid.
Individual particles in the water sample are not due to
the presence of cyanuric acid. Cyanuric acid causes
a very finely distributed cloudiness with a milky
appearance.
4.
Press TEST key, and the result is then displayed in
mg/l cyanuric acid.
The display shows the result in mg/l cyanuric acid:
32