Measuring capacity
Mains supply connection
Accuracy
Dimensions (H x W x D)
Humidity
Operating temperature range
Shut-off cycle
1.3. Charging the storage battery (A-version)
The battery is charged via an electronic circuit built into the unit. The charging process is
started when the specified power supply (see 1.2.) is connected to the charging socket,
irrespective of whether the unit is switched on or off. A new storage battery pack requires
several charging/discharging cycles to reach the specified capacity.
An empty storage battery should be charged in the unit for approx. 20 hours (max. charging
current 30 mA). Longer charging times pose no risk but may lead to a "memory effect" over
time (in other words, reduced capacity). The same applies to the charging of only partly
discharged storage batteries.
Wherever possible, therefore, the storage battery should be used until the LOW BAT indicator
lights up and then charged in the unit for approx. 20 hours.
*
PC 01
Alkali battery
Storage battery pack (600 mAh)
15 V= / 100 mA, plus polarity: pluspole inside
0.05 – 1 mg/l: ± 0.05 mg/l
Chlorine
1 – 2 mg/l: ± 0.10 mg/l
2 – 3 mg/l: ± 0.20 mg/l
3 – 4 mg/l: ± 0.30 mg/l
4 – 6 mg/l: ± 0.40 mg/l
pH value
Cyanuric acid *
In the presence of high-frequency electromagnetic fields
> 3 V/m, an additional deviation of 15 % may occur relative
to the specified measuring accuracy in the range between
230 and 280 MHz.
instrument
92 x 180 x 220 mm
case
95 x 440 x 340 mm
30 – 90 % rel. (non condensing)
0 – 40 ° C
approx. 10 minutes after last key pressed, with loss of data
Technical changes without notice!
22
approx. 8 hours
approx. 15 hours
± 0,1 pH
± 5 mg/l
approx. 0.8 kg
approx. 2.3 kg