Calibration is now done electronically
A. Switch on calibration mode
Calibration can now be done electronically using the remote control.
1. Switch off the tool and, with the
press the
-button). Keep the
- LED lights up. The
2. One of the three LEDs will light up to show which axis can be calibrated.
X-axis =
left LED, Y-axis
3. Calibration can now be continued using the remote control.
To change over between the X and Y axes, press the
4. To adjust the Z axis, end calibration for the X and Y axes (see D). Position the
tool vertically, switch on in calibration mode, and again perform the complete
calibration process.
B. Checking and calibration of the X/Y/Z axes
See page 22
C. Correcting the calibration
Using the
and
move the point from its present position to the level of reference point M or B2.
D. Finishing the calibration
Cancel: By switching off the tool, the calibration adjustments made so far will be
lost and the preceding condition restored.
Save:
By pressing the
ENGLISH
-button depressed, switch on again (briefly
-button depressed down until the
-button can now be released.
middle LED,
buttons of the remote control (by repeated pressing),
-button, the calibration is saved.
Calibration
Z-axis =
right LED
-button.
23