Hazardous area
Power supply
Display
Weight display
Weighing platform type
Sensor quantity
Weighing platform quantity
Refresh rate
Sensor excitation voltage
Minimum sensitivity
Keyboard
Communication
Metrological Approval
30491430
|
06
|
09/2020
Item
IND256x is approved for use in Zone 1/Zone 21 hazardous areas
AC power (187-253V 50/60Hz) (ATEX and IECEx version)
DC power (DC 18-30 V) (ATEX and IECEx version)
APS500/501 or external IND256x NiMH Ex battery pack (ATEX, IECEx and
cFMus version)
240 x 96 pixel white backlit LCD, 25mm high digits
Display update rate: 10 Hz
Maximum 100,000 divisions
Analog load cell
Maximum of four 350-ohm load cells (minimum 87 ohm), 2 mv/V or 3
mv/V
Supports one weighing platform
>366 Hz
4.5 VDC
0.6µV/e
26 keys; 1.5mm thick membrane keyboard
Standard interface:
The mainboard is provided with an intrinsically safe RS232
communication interface
Interface options:
Intrinsically safe 4-20 mA analog output module, with16-bit D/A
conversion and 25 Hz update rate to PLC
or WiFi communication module (only in IND256x terminals factory-
configured with WiFi module);
or Intrinsically safe active current loop
or Intrinsically safe passive current loop
Communication Protocols:
Serial port input: ASCII commands - CTPZ (Clear, Tare, Print, Zero), SICS
commands (supports SICS level 0 and level 1)
Serial port output: Toledo continuous output, command print output (5
configurable templates), SICS command and report print
Europe: OIML R76; Class III, 6000e; TC10878
Global: OIML R76; Class III, 6000e; R76-2006-A-NL1-18.27
US: Class III/IIIL, nmax=10,000; CC No.: 18-099
Canada: Class III/IIIHD, nmax=10,000; AM-6115