iv Table des matières
Tensiomètre de monitorage à distance RPM-BP100 de Welch Allyn
Directions for use
RF emissions
Class B
CISPR 11
Harmonic emissions
Not applicable
IEC 61000-3-2
Voltage fluctuations/
Not applicable
flicker emissions
IEC 61000-3-3
The Welch Allyn Remote Monitoring Blood Pressure Device RPM-BP100 is intended for use in the electromagnetic
environment specified below. The customer or the user of the Welch Allyn Remote Monitoring Blood Pressure Device
RPM-BP100 should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Electrostatic discharge
±6 kV contact
(ESD)
± 8 kV air
IEC 61000-4-2
Electrical fast
±2 kV for power supply
transient/burst
lines
IEC 61000-4-4
Surge
± 1 kV line(s) to line(s) ± 1 kV differential
IEC 61000-4-5
Voltage dips, short
<5% U
T
interruptions and
(>95% dip in U
) for 0.5
voltage variations on
T
cycle
power supply input
lines
IEC 61000-4-11
40% U
T
(60% dip in U
) for 5
T
cycles
70% U
T
(30% dip in U
) for 25
T
cycles
<5% U
T
(>95% dip in U
) for 5
T
seconds
Appendices

Electromagnetic emissions

are very low and are not likely to cause any interference in nearby
electronic equipment.
The Welch Allyn Remote Monitoring Blood Pressure Device RPM-BP100
is suitable for use in all establishments other than domestic
establishments and those directly connected to the public low-voltage
power supply network that supplies buildings used for domestic
purposes.
Electromagnetic immunity
Compliance level
Electromagnetic environment - guidance
± 6 kV contact
Floors should be wood, concrete or ceramic
tile. If floors are covered with synthetic
± 8 kV air
material, the relative humidity should be at
least 30%.
±2 kV for power supply
Mains power quality should be that of a typical
lines
commercial or hospital environment.
Mains power quality should be that of a typical
mode
commercial or hospital environment.
Mains power quality should be that of a typical
<5% U
T
commercial or hospital environment. If the user
(>95% dip in U
) for 0.5
of the Welch Allyn Remote Monitoring Blood
T
cycle
Pressure Device RPM-BP100 requires continued
operation during power mains interruptions, it
is recommended that the Welch Allyn Remote
40% U
T
Monitoring Blood Pressure Device RPM-BP100
be powered from an uninterruptible power
(60% dip in U
)for 5
T
supply or a battery.
cycles
70% U
T
(30% dip in U
) for 25
T
cycles
<5% U
T
(>95% dip in U
) for 5
T
seconds
25