6 Introductory information
CAUTION This device cannot be used with HF surgical equipment at the
same time.
CAUTION To avoid measurement errors, please avoid the condition of
strong electromagnetic field radiated interference signal or electrical fast
transient/burst signal.
CAUTION Manufacturer will make available on request circuit diagrams,
component parts list etc.
CAUTION The materials of the cuff have been tested and found to comply
with requirements of ISO 10993-5:2009 and ISO 10993-10:2010. It will not
cause any potential sensitization or irritation reaction.
CAUTION Before use, make sure the device functions safely and is in
proper working condition.
CAUTION Use the device in the environment which was provided in the
user manual. Otherwise, the performance and lifetime of the device will
been impacted and reduced.
CAUTION Do not attempt to repair the unit yourself in the event of
malfunctions. Only have repairs carried out by authorized service centers.
CAUTION Report any unexpected operation or events to the
manufacturer.
CAUTION The device does not need to be calibrated in two years of
reliable service.
CAUTION Use a soft cloth to clean the entire unit. Do not use any
abrasive or volatile cleaners.
Content list
The following items are in the device box:
•
Blood pressure monitor
•
Cuff
•
(4) AA alkaline batteries
•
Directions for use
Welch Allyn Remote Monitoring Blood Pressure Device RPM-BP100
Mode d'emploi
Indications des alarmes
Symptôme
Cause de base
Hors plage. Soit SYS > 230 mmHg ou DIA >
130 mmHg. Le symbole peut apparaître dans
la zone SYS ou DIA de l'écran.
Hors plage. Soit SYS < 60 mmHg ou DIA <
40 mmHg. Le symbole peut apparaître dans
la zone SYS ou DIA de l'écran.
Utilisation du moniteur
19
Solution
Appuyez sur le bouton
d'
alimentation
et maintenez-le
enfoncé pour rétablir la connectivité
Bluetooth.
Effectuez de nouveau une mesure. Si le
problème persiste, contactez votre
médecin.
Priorité de l'alarme = Bas !
Appuyez sur le bouton
d'
alimentation
et maintenez-le
enfoncé pour rétablir la connectivité
Bluetooth.
Effectuez de nouveau une mesure. Si le
problème persiste, contactez votre
médecin.
Priorité de l'alarme = Bas !