Alarm
Unit is over-temperature
Alarm Condition
Pause
Alarm
Therapy in-active
Alarm Condition
Pause
Alert
Mattress not connected!
Connect mattress?
Yes
Alert
Mattress not connected!
Ready to start?
Over Temperature Alarm
This screen is displayed when the therapy unit has
overheated, this may be due to a blockage or a high
ambient temperature.
Check that the air vents on the system are clear from
obstruction and that the air filter is clean.
Reset
Power the system down and allow it to
cool sufficiently.
Therapy Inactive Alarm
This screen is displayed when the system has been
left in an alert condition for more than two minutes.
The alarm sounder may be paused for two minutes
by pressing the pause button.
Reset
Mattress Not Connected Alert
This screen is displayed when the system is powered
up before connecting the air supply connector to
the therapy unit.
Connect the air supply connector and press the YES
button to clear this alert.
No
Press NO to display the CPR Alarm screen.
This screen is displayed after pressing the YES button
on the previous screen.
Connect the air supply connector and press YES to
begin therapy and clear this alert.
Yes
28