IRREGULAR PULSE RHYTHM SYMBOL
Pulse rhythm can be disturbed from talking,
moving or arrhythmias. This product displays
[
] when the difference of shortest and
longest time intervals between pulse waves
is over 25%, indicating irregular pulse
rhythm.

MEMORY FUNCTION

The measured values are automatically saved for later review in either one of two memory
banks, [1] or [2]. These two memory banks can be used to save readings of two persons
separately or to save morning readings and evening readings separately.
Each bank can save up to 60 readings. When the number of stored readings reaches 60,
the oldest reading will be deleted to record new reading.
Saved readings are recalled with measurement date and time when clock function is
activated. *The clock needs to be activated before measurement for date and time to be
saved with the measured valued.
[Err] results are not saved.

CARE AND MAINTENANCE

Keep the product clean. Inspect its cleanliness after use. To clean, use only a soft dry
cloth. Do not use gasoline, paint thinner, or other strong solvents. Since the cuff may
absorb perspiration and other fluids, inspect it for stain and discoloration after each use.
When cleaning the cuff, use a synthetic detergent and gently rub the surface. Air dry
thoroughly. Make sure fluid never gets in AIR HOSE. Do not machine wash or scrub it.
When storing the product, do not place heavy objects on it. Do not fold or bend the cuff
forcibly. To disconnect the cuff from the main unit, do not pull AIR PLUG but hold AIR
PLUG and unplug it. Do not coil AIR HOSE too tightly. When the product has been stored
at a temperature below the freezing point, keep it for at least 1 hour in a warm place before
using. Remove the batteries if the product is to be stored for an extended period of time.
Keep the batteries out of reach of children.
regular pulse rhythm
An example of irregular pulse rhythm
compared to regular pulse rhythm
We suggest that you have your blood pressure monitor checked every 2 years. This
operation may only be performed by the manufacturer or by firms authorized by the
manufacturer.

PRODUCT SPECIFICATIONS

Model
: DSK-1031
Operating Principle
: Oscillometric method
Indicator
: 15 digits liquid crystal display
Pressure Indicating Range
: 3 to 300 mmHg (cuff pressure)
Measuring Range
: 50 to 250 mmHg (systolic), 40 to 180 mmHg (diastolic),
40 to 160 bpm (pulse rate)
Accuracy*
: ±3 mmHg (cuff pressure), ±5% of reading (pulse rate)
Inflation
: Automatic inflation (FUZZY)
Deflation
: Automatic (electric control valve)
Exhaust
: Automatic exhaust valve
Power Supply
: Four 1.5 volt LR6 (AA alkaline) batteries or AC adaptor
ADP-W5 series
Power Consumption
: 4W (max.)
Electric Rating
: DC6V/4W, with four LR6 batteries
Adaptor; AC100-240V, 50-60Hz, 0.12A,
Unit; DC6V,
Memory
: 2 banks, each stores 60 readings
Operating Condition
: +10˚C to +40˚C, 30% to 85% RH (noncondensing)
Transportation/Storage Condition
: -20˚C to +60˚C, 10% to 95% RH (noncondensing)
Cuff
: Model; DSC-EP10, Coverage arm circumference; 22 to
42 cm, applied part; BF
Main Unit
: Weight; Approx. 250 g (without batteries)
Size; Approx. 115 x 115 x 65.9 mm (W x D x H)
Protection against harmful ingress of water
: IPX0 (IEC 60601-1)
-4-
500mA, with designated AC adaptor