Technical Description - Nissei DSK-1031 Instructions

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 27
T E C H N I C A L D E S C R I P T I O N
DSK-1031 complies with the EMC, electromagnetic compatibility, standard, IEC60601-1-2. Refer to the following tables for specific information regarding
compliance to the standard.
DSK-1031, as a medical electrical equipment, needs special precautions regarding EMC and needs to be installed and put into service according to the EMC
information provided below.
• Portable and mobile RF communications equipments can affect the device.
• The use of accessories other than those specified in this manual may result in increased emissions or decreased immunity of the device.
• DSK-1031 should not be used adjacent to or stacked with other equipment.
Guidance and manufacturer's declaration - electromagnetic emissions -
DSK-1031 is intended for use in the electromagnetic environment specified below. The customer or the user of DSK-1031 should assure that it is used in
such an environment.
Emissions test
Compliance
RF emissions CISPR 11
Group 1
RF emissions CISPR 11
Class B
Harmonic emissions IEC 61000-3-2
Class A
Voltage fluctuations/flicker emissions IEC 61000-3-3
Complies
Guidance and manufacturer's declaration - electromagnetic immunity -
DSK-1031 is intended for use in the electromagnetic environment specified below. The customer or the user of DSK-1031 should assure that it is used in
such an environment.
Immunity test
IEC 60601 test level
Electrostatic discharge (ESD)
±6 kV contact
IEC 61000-4-2
±8 kV air
Electrical fast transient/burst
±2 kV for power supply lines
IEC 61000-4-4
±1 kV for input/output lines
Surge IEC 61000-4-5
±1 kV differential mode
±2 kV common mode
<5% U
(>95% dip in U
) for 0,5
T
T
Voltage dips, short
cycle
interruptions and voltage
40% U
(60% dip in U
) for 5
T
T
variations on power supply
cycles
input lines IEC 61000-4-11
70% U
(30% dip in U
) for 25
T
T
cycles
<5% U
(>95% dip in U
) for 5
T
T
sec
Power frequency (50/ 60 Hz)
3 A/m
magnetic field IEC 61000-4-8
NOTE U
is the a.c. mains voltage prior to application of the test level.
T
Electromagnetic environment - guidance
DSK-1031 uses RF energy only for its internal function. Therefore, its RF emissions are
very low and are not likely to cause any interference in nearby electronic equipment.
DSK-1031 is suitable for use in all establishments, including domestic establishments
and those directly connected to the public low-voltage power supply network that
supplies buildings used for domestic purposes.
Compliance level
Electromagnetic environment - guidance
±6 kV contact
Floors should be wood, concrete or ceramic tile. If
±8 kV air
floors are covered with synthetic material, the
relative humidity should be at least 30 %.
±2 kV for power supply lines
Mains power quality should be that of a typical
±1 kV for input/output lines
commercial or hospital environment.
±1 kV differential mode
Mains power quality should be that of a typical
±2 kV common mode
commercial or hospital environment.
<5% U
(>95% dip in U
) for 0,5
Mains power quality should be that of a typical
T
T
cycle
commercial or hospital environment. If the user of
40% U
(60% dip in U
) for 5
DSK-1031 requires continued operation during
T
T
cycles
power mains interruptions, it is recommended that
70% U
(30% dip in U
) for 25
DSK-1031 is be powered from an uninterruptible
T
T
cycles
power supply or a battery.
<5% U
(>95% dip in U
) for 5
T
T
sec
3 A/m
Power frequency magnetic fields should be at
levels characteristic of a typical location in a
typical commercial or hospital environment.
Guidance and manufacturer's declaration - electromagnetic immunity -
DSK-1031 is intended for use in the electromagnetic environment specified below. The customer or the user of DSK-1031 should assure that it is used in
such an environment.
Immunity test
IEC 60601 test
Compliance
level
level
Conducted RF
3 Vrms, 150 kHz
3 V
IEC 61000-4-6
to 80 MHz
Radiated RF IEC
3 V/m, 80 MHz to
3 V/m
61000-4-3
2.5 GHz
NOTE1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from
structures, objects and people.
Field strength from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM
a
radios broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF
transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which DSK-1031 is used exceeds the
applicable RF compliance level above, DSK-1031 should be observed to verify normal operation. If abnormal performance is observed, additional
measures may be necessary, such as reorienting or relocating DSK-1031.
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than [V1] V/m.
b
Table 206 - Recommended separation distances between portable and mobile RF communications equipment and DSK-1031 -
DSK-1031 is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of DSK-1031
can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment
(transmitters) and DSK-1031 as recommended below, according to the maximum output power of the communications equipment.
Separation distance according to frequency of transmitter, m
Rated maximum output
power of transmitter, W
150 kHz to 80 MHz, d=[3.5/V1]√P
0.01
0.12
0.1
0.38
1
1.2
10
3.8
100
12
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in metres (m) can be estimated using the
equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the
transmitter manufacturer.
NOTE1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and
people.
-13-
Electromagnetic environment - guidance
Portable and mobile RF communications equipment should be used no closer to any part of
DSK-1031, including cables, than the recommended separation distance calculated from the
equation applicable to the frequency of the transmitter.
Recommended separation distance
d=[3.5/V1]√P
d=[3.5/E1]√P, 80MHz-800MHz
d=[7/E1]√P, 800MHz-2.5GHz
Where P is the maximum output power rating of the transmitter in watts (W) according to the
transmitter manufacturer and d is the recommended separation distance in metres (m).
Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey,
should be less than the compliance level in each frequency range.
b
Interference may occur in the vicinity of equipment marked with the following symbol:
80 MHz to 800 MHz, d=[3.5/E1]√P
800 MHz to 2.5 GHz, d=[7/V1]√P
0.12
0.38
1.2
3.8
12
a
0.23
0.73
2.3
7.3
23

Publicité

Table des Matières
loading

Table des Matières