Technical Information
Description
N.
Read/ Write
106
R/W
107
R/W
108
R/W
109
W
110
R/W
111
R/W
120
R/W
122
R/W
125
R/W
126
R/W
182/192
Data
Description
Type
L
Water cooling (write only in stop)
L
Active Stop (write only in stop)
N
Baud rate
L
Pump life/ cycle time/ cycle number reset
L
Interlock type
(default = 1)
L
Analog output type: output voltage signal
proportional to frequency or power
N
Rotational frequency setting (Hz)
L
Set vent valve on/off
L
Set the vent valve
operation
N
Vent valve opening
delay
(expressed in 0.2 sec)
TwisTorr 74 FS AG On-board Controller User Manual / 87-901-054-01
Permitted Values
0 = NO
1 = YES
0 = NO
1 = YES
(default = 0)
600 = 0
1200 = 1
2400 = 2
4800 = 3
9600 = 4
(default = 4)
To reset, write '1'
Impulse = 0
Continuous = 1
0 = frequency
1 = power
(default = 0)
2 = Pump temperature
3 = Pressure
1100 to 1167 (default
= 1167)
0 = Closed
1 = Open
(default = 1)
Automatic = 0
(see note 1.)
On command = 1
(see note 2.)
(default=0)
0 to 65535
(corresponding to 0
to 13107 sec)
(default = 15)