Télécharger Imprimer la page

Agilent Technologies TwisTorr 74 FS AG Mode D'emploi page 166

Masquer les pouces Voir aussi pour TwisTorr 74 FS AG:

Publicité

Technical Information
J1 – IN-OUT
For example:
Figure 20
It is possible to delay the set point checking for a programmable
delay time.
The PROGRAMMABLE SET POINT signal has the following default
settings:
The A-PLUS (optional) allows the operator to set all the programmable features.
NOTE
166/192
reference quantity: frequency
threshold: 867 Hz
hysteresis: 2 %
activation type: "high level"
reference quantity: frequency
threshold: 867 Hz
hysteresis: 2 %
activation type: high level
delay time: 0 second
TwisTorr 74 FS AG On-board Controller User Manual / 87-901-054-01

Publicité

loading

Ce manuel est également adapté pour:

X3509-64030X3509-64021