WIKA Cryo Gauge 712.15.160 Mode D'emploi page 11

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

8. Transmitter for level measurement
Specifications
Power supply U
B
Influence of power supply
Permissible residual ripple ≤ 10 % ss
Output signal
Permissible max. load R
Effect of load
Adjustability
Zero point, electrical
Scale selection
Linear error
Permissible ambient tem-
perature range
Compensated temperature
range
Temperature coefficients in the compensated temperature range
Mean TC zero point
Mean TC span
Safety-related maximum
values
Power supply U
i
Short-circuit current I
Power P
i
Internal capacitance C
Internal inductance L
Medium temperature
Ambient temperature
Electrical connection
Wiring protection
Ingress protection
Wiring details, 2-wire
1) Only possible within 30 seconds of connecting the supply voltage
WIKA operating instructions differential pressure gauge models 712.15.160, 732.15.160
Models 891.44 and 892.44 (Ex version)
DC 12 < U
≤ 30 V (≥ 14 V with Ex version)
B
≤ 0.1 % of full scale/10 V
4 ... 20 mA, 2-wire
for non-Ex versions, model 891.44:
A
R
≤ (U
- 12 V) / 0.02 A with R
A
B
for Ex versions, model 892.44:
R
≤ (U
- 14 V) / 0.02 A with R
A
B
≤ 0.1 % of full scale value
Adjustment of the zero point through brief bridging of terminals 5 and 6
or using the "scale selection switch" option, selectable via button
4 scales selectable via BCD switch
≤ 1.0 % of span (terminal method)
-40 ... +80 °C, -40 ... +60 °C with oxygen
-40 ... +80 °C
≤ 0.3 % of span/10 K
≤ 0.3 % of span/10 K
Ex version
DC 14 ... 30 V
≤ 100 mA
i
≤ 1 W
12 nF
i
negligible
i
-40 ... +80 °C, -40 ... +60 °C with oxygen
-40 ... +60 °C (T6)
Angular connector, 180° rotatable, wire protection, cable gland M20 x 1.5,
incl. strain relief, connection cable: Outer diameter 7 ... 13 mm, conductor
cross-section 0.14 ... 1.5 mm
Protection against reverse polarity and overvoltage
IP65 per EN/IEC 60529
Ground, connected
to case 2)
U
+/Sig
B
+0 V/Sig-
in Ohm and U
in Volt
A
B
in Ohm and U
in Volt
A
B
, temperature resistance up to 60 °C
2
Terminals 3, 4, 5 and 6: for internal use only
2) This connection must not be used
for equipotential bonding. The
instrument must be incorporated in the
equipotential bonding via the process
connection.
EN
1)
11

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Cryo gauge 732.15.160

Table des Matières