1. Error messages
Error
Possible cause
Display Err - 300
Excessive cuff
pressure. The cuff
was inflated to its
maximum due to the
movement of the
arm or body.
The air hose may be
kinked or the air plug
blocked.
Display Err - 1
Measurement error;
measurement could
not be carried out
due to movement or
speaking.
The acquired values
lie outside the mea-
suring range of the
device (see „Technical
data" on page 62).
Display Err - 2
Inflation error;
pressure built up too
quickly or too slowly.
Fault has been iden-
tified. The cuff is not
correctly attached
Display Err - 3
Deflation rate too
high or too low.
Display 0
Start/Stop button
0
was accidentally
pressed while the
batteries were being
replaced or while the
mains adapter was
being plugged in.
EN-60
Technical information
Remedy
Repeat the measurement
Do not move your arm
Do not speak
Inspect the hose line and
air plug for blockages
- Repeat the measurement
- Do not move your arm
- Do not speak
Repeat the measurement
Check that the air connec-
tor is fitted properly, repeat
measurement.
- Repeat the measurement
- Do not move your armDo
not speak
Call the Service hotline if
this occurs repeatedly.
Switch unit off and then on
again using the Start/Stop
button.
Technical information
Error
Possible cause
Arrhythmia, mo-
vement disorders,
trembling, wobbling
(objects), deep bre-
athing etc.
Atrial fibrillation
The motion control
detected movement
during the measure-
ment.
Measured values/
Too little rest before
pulse values are
taking the measure-
extremely high or
ment. Movement or
low.
talking during the
measurement.
Incorrect measure-
ment posture
Measurements
Blood pressure is not
taken in direct
a fixed value. It can
succession produce
go up or down by
different measured
more than 20 mmHg
values.
within a few minutes.
Measurement was
Device detects a fault
interrupted, the
or that the diastolic
cuff deflated to 0
value is too low.
and the measure-
Movement during
ment restarted.
measurement
No display after de-
Batteries inserted
vice is switched on
incorrectly
Batteries flat
Battery contacts
dirty.
E
Remedy
Repeat measurement after
resting for 3-5 minutes.
Listen out for the device's
pulse beep signals. Possible
influences on the measure-
ment results (depending on
the severity of the arrhyth-
mia) must be discussed
with your doctor!
- Repeat the measurement
- Do not move your arm
- Do not speak
Repeat the measure-
ment after a break of 3-5
minutes. Pay attention to
the usage information on
page 45.
Repeat the measurement
and maintain the correct
posture. (Page 49)
In order to obtain compara-
ble results, always carry out
the measurements under
the same conditions.
Device shows no fault.
Repeat measurement
If necessary, interrupt
measurement and repeat
after 5 minutes' rest.
Check position of batteries.
Replace batteries.
Clean battery contacts with
a dry cloth.