11. Accessories and replacement parts
Accessories and replacement parts are available from the corresponding service address (according
to the service address list). Please state the corresponding order number.
Designation
Universal cuff (23-43 cm)
12. What if there are problems?
Problem /
Error Message
The blood pressure meas-
urement result shows
The blood pressure meas-
urement result shows
The blood pressure meas-
urement result shows
/
The blood pressure meas-
urement result shows
The ECG measurement
flashes on the display.
Item number and/or order number
164.226
Possible cause
1. Inflating too slow or can-
not inflate normally.
2. Air leakage of Cuff
Abnormality found in air
releasing of arm cuff.
Abnormality found in blood
pressure readings.
Low battery.
The ECG signal is unstable
or too weak.
52
Solution
Please repeat the measurement.
Make sure that the cuff tube is
properly inserted and that you
do not move or talk. Re-insert
the batteries if necessary, or
else replace them.
Repeat the measurement as per
the instructions in these Instruc-
tions for use.