Table 2 ‐ IMMUNITY TEST LEVELS
Phenomenon
Electrostatic discharge
Radiated RF electroma-
gnetic fields
Proximity fields from RF
wireless communications
equipment
Rated power frequency
magnetic fields
Table 3 - Test specifications for ENCLOSURE PORT IMMUNITY to RF wireless communi-
cations device
Test
Band
frequency
(MHz)
(MHz)
385
380 to 390
450
430 to 470
710
704 to 787
745
780
Basic EMC standard
IEC 61000‐4‐2
IEC 61000‐4‐3
IEC 61000‐4‐3
IEC 61000‐4‐8
Service
Modulation Maximum
TETRA 400
Pulse
modulation
18 Hz
GMRS 460,
FM
FRS 460
±5 kHz
deviation
1 kHz sine
LTE Band
Pulse
13, 17
modulation
217 Hz
230
IMMUNITY TEST LEVELS
±8 kV contact
±2 kV, ±4 kV, ±8 kV, ±15 kV air for en-
closure port
3 V/m
80 MHz to 2.7 GHz
80 % AM at 1 kHz for enclosure port
See table 3
30 A/m
50 Hz and 60Hz for enclosure port
Distance
power (W)
(m)
1.8
0.3
2
0.3
0.2
0.3
IMMUNITY
TEST LEVEL
(V/m)
27
28
9