Figure 8 Analog output connections
Recorder wires
Output 2-
Output 2+
Output 1-
Output 1+
1. Open the controller cover.
2. Feed the wires through the cable gland.
3. Adjust the wire as necessary and tighten the cable gland.
4. Make connections with twisted-pair shielded wire and connect the shield at the controlled
component end or at the control loop end.
• Do not connect the shield at both ends of the cable.
• Use of non-shielded cable may result in radio frequency emission or susceptibility levels higher
than allowed.
• Maximum loop resistance is 500 ohm.
5. Close the controller cover and tighten the cover screws.
6. Configure outputs in the controller.
4.10 Discrete input wiring connections
Potential Electrocution Hazard. Always disconnect power to the instrument when making electrical
connections.
Potential electrocution hazard. In order to maintain the NEMA/IP environmental ratings of the
enclosure, use only conduit fittings and cable glands rated for at least NEMA 4X/IP66 to route
cables in to the instrument.
Table 3 Output connections
W A R N I N G
W A R N I N G
Circuit board position
4
3
2
1
English 19