y
9. Do not throw the battery in a fire, it may explode.
10. Electromagnetic interference – Do not use your doppler in the immediate vicinity of your
en
mobile phone or any radio transmitter; this could cause electromagnetic interference.
11. This device cannot be used with a defibrillator or any high-frequency medical device.
12. Avoid shocks and knocks to the device.
13. Only use your doppler and its accessories for the functions it is intended for.
14. Do not let children use your device and its accessories.
15. The doppler should be applied to healthy skin. If you are not sure of the physiological state
of the skin, do not use it.
16. The recommended gel is a hypoallergenic product, non-irritating and non-corrosive. If you
er,
make a skin reaction with the gel, stop the application and consult a physician.
-
17. The doppler should be cleaned after each use according to recommendations of « Mainte-
ade
nance » section.
GY.
18. The doppler should not be used before the 12th week of pregnancy.
19. At the end of its life cycle, the doppler and its accessories should be recycled in accor-
dance with the provisions in force.
20. Be careful to strangulation due to cables, particularly due to excessive length.
07. TECHNICAL FEATURES
Trademark
Commercial reference
Safety Class
Electromagnetic
compatibility (EMC)
Internal power supply
Appropriate use
Environment
e
e
d
Physical features
Probe
ust
Babydoppler Connect
Babydoppler Connect
Type CF applied part
IEC 60601-1:2005 + A1:2012
IEC 60601-1-2 Class B group 1
This device operates on 3 V DC
(2 x 1.5V AAA size batteries)
Power consumption < 1 W
As from the 12
week of pregnancy
th
Working
•Temperature : between +5°C and 40°C
•Humidity ≤ 80%
•Atmospheric pressure: between 70 kPa and 106 kPa
Transport and storage
•Temperature: between -10°C and 55°C
•Humidity < 93%
•Atmospheric pressure: between 50 kPa à 106 kPa
Dimensions : 105 x 90 x 36 mm
Mass : around 80 g (including the batteries)
Nomimal frequency : 3 MHz
Operating frequency : 3 MHz ± 10%
Output acoustic power : < 1 MPa
I
: < 100 mW/cm
2
spta
MI et TI < 1
Output intensity of ultrasound : I
Effective Radiating Area of Transducer <208mm²
< 20 mW/cm
2
ob
11