Technical specifications
Pressure–flow
curve
60
50
40
30
20
10
0
4
6
8
Mask pressure (cm H
Dead space
information
Therapy pressure
Resistance
Environmental
conditions
Gross dimensions
Mask setting
options
The mask contains passive venting to protect against
rebreathing. As a result of manufacturing variations,
the vent flow rate may vary.
10
12
14
0)
2
Physical dead space is the empty volume of the mask
to the end of the swivel. Using the large cushions it
is 106 mL.
4 to 20 cm H
O
2
Drop in pressure measured (nominal)
at 50 L/min: 0.4 cm H
at 100 L/min: 1.4 cm H
Operating temperature: +41°F to 104°F
(+5°C to +40°C)
Operating humidity: 15% to 95% relative humidity
non-condensing
Storage and transport temperature: -4°F to 140°F
( -20°C to +60°C)
Storage and transport humidity: up to 95% relative
humidity non-condensing
L:
16.34" (H) x 3.70" (W) x 1.77" (D)
(415 mm (H) x 94 mm (W) x 45mm (D))
Mask fully assembled with short tube assembly – no
headgear.
For S9: Select 'Pillows'
For other devices: Select 'SWIFT', otherwise select
'MIRAGE' as the mask option.
Pressure
(cm H
4
8
12
16
16
18
20
20
O
2
O
2
5
Flow
)
O
(L/min)
2
20
29
37
43
49