Selecting a Calibration Mode (pH)
Calibration is used to adapt the device to the individual sensor characteristics,
namely asymmetry potential and slope.
Access to calibration can be protected with a passcode (SERVICE menu).
First, you open the calibration menu and select the calibration mode:
CAL_PH
Depending on confi guation setting:
AUTO
MAN
DAT
CAL_ORP
ORP calibration
P_CAL
Product calibration (calibration with sampling)
ISFET-ZERO
Zero adjustment. Required for ISFET sensors. Subsequently you
can conduct either a one or a two-point calibration.
CAL-RTD
Temperature probe adjustment
To preset CAL_PH (CONF menu / confi guration):
1) Hold meas key depressed (> 2 s) (measuring mode)
2) Press menu key: the selection menu appears
3) Select CONF mode using left / right arrow key
4) Select "SENSOR" – "CALMODE": AUTO, MAN, DAT.
Press enter to confi rm.
1
Automatic buff er recognition (Calimatic)
Manual buff er input
Input of premeasured electrode data
Selection menu
Selected mode
(blinks)
4
3
2
21