4
Installation Instructions
• Installation of the device must be carried out by trained experts in accordance
with this user manual and as per applicable local and national codes.
• Be sure to observe the technical specifi cations and input ratings during
installation!
• Be sure not to notch the conductor when stripping the insulation!
• Before connecting the device to the power supply, make sure that its voltage lies
within the range 80 to 230 V AC/DC or 24 to 60 V DC.
• A signal current supplied to the current input must be galvanically isolated.
If not, connect an isolator module.
• All parameters must be set by a system administrator prior to commissioning.
Terminals:
suitable for single or stranded wires up to 2.5 mm
Application in Hazardous Locations:
For use in hazardous locations, see separate "Certifi cates" document:
• IECEx
• ATEX
Installation
2
(AWG 14)