Selecting the Calibration Mode (pH)
5
Menu item
Temp detection during
calibration
(Manual temperature)
Calibration Mode
(AUTO: Buff er set)
Action
Select mode using keys:
AUTO: Measured by sensor
MAN: Direct input of tempera-
ture, no measurement (see next
step)
EXT: Temperature specified
via current input (only if TAN E
enabled)
Press enter to confirm.
Modify digit using keys,
select next digit using
keys.
Press enter to confirm.
Select CALMODE using
keys:
AUTO: Calibration with Calimatic
buffer set recognition
MAN: Manual entry of buffer
solutions
DAT: Input of adjustment data of
premeasured sensors
Press enter to confirm.
Select buffer set using
keys (see buffer tables for nom-
inal values)
Press enter to confirm.
Choices
-00-...-10-,
(-U1-, see Appendix)
Pressing the info key displays
the manufacturer and nominal
values in the lower line.
19