sense of EN60601-1, which would present an unacceptable risk to patients, operators or
third parties should the power supply fail or malfunction.
Warning:
Portable RF communications equipment (radios) including accessories, such as antenna
cables and external antennas, should not be used in closer proximity than 30 cm (12 in-
ches) to parts and cables of the ri-magic®HPLED specified by the manufacturer. Failure to
comply may result in reduced device performance.
Guidelines and manufacturer's declaration – electromagnetic emissions
The ri-magic HPLED examination lamp is intended for use in an electromagnetic en-
vironment as specified below. The customer or user of the headlight should ensure that
it is used in such an environment.
Emissions test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emis-
sions
IEC 61000-3-2
Voltage fluctu-
ations/flicker
emissions
IEC 61000-3-3
Compliance
Groupe 1
Class A
Class A
Compliant
Electromagnetic environment - guidance
The ri-magic® HPLED examination lamp
only uses RF energy for its internal functi-
oning. Therefore, its HF emissions are very
low and are unlikely to cause interference
in nearby electronic devices.
The ri-magic® HPLED examination lamp
is suitable for use in all establishments,
including domestic establishments and
those directly connected to the public
low-voltage network that supplies buildings
for domestic purposes.